Flow Theories

When you hear the term mass flow rate, you usually think of units such as kg/h or g/min and the like. However, if you look at practice, you realize that working with volume units is predominant. Now, all that had to be done was to agree on standard conditions for converting mass into volume.

Standard conditions (ln/min): The standard conditions were defined as 0 °C and a pressure of 1013.25 mbar absolute. The relevant volume units are identified by the subscript letter n: ln/min.

Standard conditions (ls/min): Another specification for converting mass to volume is the standard conditions, which are based on 20 °C instead of 0 °C and are marked accordingly with an s instead of the n: ls/min.

Note: Internationally, the terms sccm (standard cubic centimeters per minute) and slm (standard liter per minute) are often used; the same definition applies here as for mln/min (0 °C, 1013.25 mbar abs).

Note: If the difference between ln/min and ls/min is not taken into account, this results in an error of approx. 7%.

You are welcome to give us your reference points, your instrument will then be configured accordingly.

These abbreviations of the units of mass flow beginning with an “S” refer to the usual reference conditions: 1013.25 hPa(abs) and 0 °C.

Note: Reference temperature and pressure conditions of the flow unit of the instruments are always listed on the calibration certificate.

The turndown ratio of an instrument indicates the range in which a flow meter or controller can accurately measure a liquid. In other words, it corresponds to the upper limit of a measuring range compared to the lower limit. This is expressed as a ratio and calculated using a simple formula:

Measuring range = maximum / minimum flow rate

For example, if a particular flow meter has a measuring span of 50:1, it can measure the maximum flow to within 1/50. Assuming a flow meter has a full-scale value of 20 ln/min, the flow meter will measure the flow to the end of the range of up to 0.4 ln/min. Bear in mind that the maximum or minimum possible flow rate of a measuring device or controller can have a larger measuring range than the adjustable measurable and controllable range. For example, it would be possible for a mass flow controller to have a measuring range from a minimum of 0.16 ln/min to a maximum of 25 ln/min, but the actual measuring range is determined by the reduction ratio. In relation to this example, this means If the calibrated high flow rate is 25 ln/min, the lowest measurable value corresponds to 0.5 ln/min (1/50 of 25). If the application requires the calibrated minimum flow to be 0.1 ln/min, then the measurable maximum flow is 5 ln/min (50 times 0.1).

Bronkhorst has developed a database in which more than 1800 fluid data are stored. Using the application software, various calculations can be performed online on our website www.fluidat.com. We invite our customers to use this website after a free registration. After receiving your login details, you can calculate conversion factors and tables for Bronkhorst thermal gas and liquid meters, Kv values and orifice sizes of Bronkhorst control valves, pressure drop of Coriolis flow meters and inlet filters and perform calculations that are important for your CEM evaporation system.

FAQ

Electrics

There is a wide variety of possibilities for electrical connection and communication. All Bronkhorst mass flow/pressure meters and controllers can generally be supplied with 15 or 24 Vdc. For analog inputs and outputs, 0-5 Vdc, 0-10 Vdc, 0-20 mA and 4-20 mA are optionally available. We can also offer other options on request.

The digital communication of most product lines includes RS232, PROFIBUS DP, DeviceNet™, Modbus-RTU/ASCII, EtherCAT®, PROFINET, CANopen, Modbus TCP, EtherNet/IP, Powerlink and FLOW-BUS (Bronkhorst® fieldbus).

We offer a variety of options for fieldbus interfaces that are integrated into the flow/pressure meter and controller. Typically, the additional current consumption for the interfaces is < 75 mA for instruments used at 15 V supply voltage and < 50 mA for instruments used at 24 V supply voltage.

Additional power consumption of onboard fieldbus interfaces

In the table you can see the exact values for each fieldbus type.

Fieldbus
with 15 V supply
with 24 V supply
PROFIBUS DP
53 mA
30 mA
PROFINET
76 mA
48 mA
EtherCAT®
66 mA
41 mA
CANopen®
n.a.
48 mA
DeviceNet ™
n.a.
48 mA
EtherNet/IP
51 mA
35 mA
Modbus-TCP
51 mA
35 mA
POWERLINK
51 mA
35 mA
Questions about measurement technology in general

Basics

Unlike mass-related units such as the gram, a measure of volume is always dependent on its ambient conditions. The volume changes when the temperature and/or pressure changes. If the temperature is increased, the space occupied by the medium increases. If the pressure increases, it becomes smaller. Consequently, when specifying a volume, the reference temperature and the reference pressure must always be specified.

If I enclose a liter of air at 0 °C and an ambient pressure of 1013.25 mBar, this is referred to as a standard liter in measurement technology. If the temperature is increased to 20 °C, the gas expands by 7.3 %. This is called a standard liter. Each industry sets its own standard.

Contact

If you measure the volumetric flow rate, this refers to the volume of a medium that moves per unit of time. The corresponding units are m3/h, m3/min, CFM or ACFM. If the medium is a gas, its volume expands depending on the temperature and pressure conditions. This must be taken into account when measuring the volume. However, the weight of a gas, i.e. its mass, does not change. The mass flow rate is the measure of the movement of a mass per unit of time. The units for mass flow are kg/h and lb/min.

The clearest reference pressure for pressure measurement is absolute zero (vacuum = 0 bar), which prevails in the vacuum of the universe. A pressure that is related to this reference pressure is called absolute pressure. With relative pressure, the reference pressure is not the zero point, but the actual ambient pressure (atmospheric pressure = 1 bar). A relative deviation of the air pressure from the ambient air pressure is called gauge pressure.

FAQ

Installation

What is FLOW-BUS?

FLOW-BUS is a fieldbus developed by Bronkhorst based on RS485 technology for digital communication between multiple digital devices. This communication protocol offers the possibility of host control by a PC.

What is FlowSuite?

FlowSuite is a Windows PC application for monitoring and servicing Bronkhorst digital instruments via a single interface in a multi-device window. It provides a good insight into the dynamic behavior of the flow meters and controllers.

How to connect FLOW-BUS and FlowSuite in practice?

Bronkhorst instruments are equipped with a micro-controller to exchange parameter value information with other devices connected to the same FLOW-BUS system or via RS232 to a PC.

Here are some examples of how to correctly install Bronkhorst instruments using FLOW-BUS with FlowSuite:

  • Bronkhorst instruments are connected via FLOW-BUS, RS485 (red cable).
  • The USB port of the PC is connected to the first instrument via RS232 (green cable).
  • Bronkhorst FlowSuite™ displays the PV (process variables), controls the SP (set point), together with graphing and data logging for up to 4 graphs. The number of instruments to be connected depends on the performance of your hardware.
  • A standard RJ45 shielded (F)TP patch cable or an M12 cable supplies the FLOW-BUS signal and the 24 Vdc voltage.
  • 1A @ 24Vdc or 24 watts is the maximum power for RJ45 Shielded (F)TP patch cables, 4A for the M12 cables.

Digital multi-bus instruments (operated via the analog interface) are pin-compatible with Bronkhorst analog instruments.

Note: If you have previously set the instrument to a setpoint via a potentiometer, please let us know when you make your request. We can offer you an adapter for this operating mode. The same applies to instruments with circuit board coding HC, HD, FC and FD.

The measured values are always available simultaneously as analogue and RS-232 signals. The setpoint specification can be either analogue or via RS-232. This must be configured beforehand.

Gas properties vary according to temperature and pressure fluctuations. The EL-FLOW Prestige thermal models use the actual measurement temperature (and pressure, if applicable) for the on-board calculation of fluid properties in real time. For this reason, these devices have an embedded database in which gas properties are stored (“Fluidat-On-Board”).

The following gases are included:

EL-FLOW® Prestige models built until Dec. 2018
EL-FLOW® Prestige models built from Jan. 2019
formula
designation
formula
designation
formula
designation
Air
Air
Air
Air
CH3CI
Chloromethan
AR
Argon
AR
Air
CH3F
Fluoromethan
C2F6
Freon-116
AsH3
Arsin (Arsan)
CH4
Methan
C2H2
Acetylen (Ethin)
B2H6
Diboran
CH4S
Methanthiol
C2H4
Ethen
BCI3
Bortrichlorid
CH5N
Methylamin (Aminomethan)
C2H6
Ethan
BF3
Bortrifluorid
CHCl2F
Dichlorofluoromethan
C3H6 #2
Propen
C2CI2F4 #2
Freon-114
CHClF2
Chlorodifluoromethan
C3H8
Propan
C2Cl3F3
Freon-113
CHF3
Freon-23
CH4
Methan
C2ClF5
Freon-115
Cl2
Chlor
Cl2
Chlor
C2F4
Perfluoroethen
ClCN
Cyanchlorid
CO
Carbon monoxide
C2F6
Freon-116
ClF3
Chlortrifluorid
CO2
Carbon dioxide
C2H2
Acetylen (Ethin)
CO
Carbon monoxide
H2
Hydrogen
C2H2F2 #1
Freon-1132A
CO2
Carbon dioxide
H2S
Hydrogen sulphide
C2H3Br
Vinylbromid
COCl2
Carbonylchlorid
He
Helium
C2H3Cl
Chloroethen
COF2
Carbonylfluorid
Kr
Krypton
C2H3F
Fluoroethen
COS
Carbonylsulfid
N2
Nitrogen
C2H4
Ethen
CS2
Carbon disulphide
N2O
Nitrous oxide (laughing gas)
C2H4O #2
Epoxyethan
D2 #1
Deuterium
NF3
Nitrogen trifluoride
C2H5Cl
Chloroethan
F2
Fluorine
NH3
Ammonia
C2H6
Ethan
GeH4
German
NO
Nitric oxide
C2H6O #1
Dimethylether
H2
Hydrogen
O2
Oxygen
C2H7N #2
Dimethylamin
H2S
Hydrogen sulphide
SF6
Sulphur hexafluoride
C2H7N #3
Monoethylamin
H2Se
Hydrogen selenide
SiH4
Silan
C2N2
Dicyan (Ethandinitril)
HBr
Hydrogen bromide
C3F8
Perfluoropropan
HCl
Hydrogen chloride
C3H4 #1
Allen
HCN
Hydrogen cyanide
C3H4 #2
Methylacetylen
He
Helium
C3H6 #1
Cyclopropan
HF
Hydrogen fluoride
C3H6 #2
Propen
HI
Hydrogen iodide
C3H8
Propan
Kr
Krypton
C3H9N #3
Trimethylamin
MoF6
Molybdenum hexafluoride
C4F8
Freon-C318
N2
Nitrogen
C4H10 #1
n-Butan
N2O
Nitrous oxide (laughing gas)
C4H10 #2
Isobutan
Ne
Neon
C4H6 #3
1,3-Butadien
NF3
Nitrogen trifluoride
C4H6 #4
1-Butin
NH3
Ammonics
C4H8 #1
Cyclobutan
NO
Nitric oxide
C4H8 #2
1-Buten
O2
Oxygen
C4H8 #3
Buten (2-) (cis)
OF2
Oxygen difluoride
C4H8 #4
Buten (2-) (trans)
PH3
Phosphin
C4H8 #5
2-Methylpropen
SF4
Sulphur tetrafluoride
C5H12 #2
2,2-Dimethylpropan
SF6
Sulphur hexafluoride
C5H12 #3
n-Pentan
Si2H6
Disilan
CBr2F2
Dichlorodifluoromethane
SiH2Cl2
Trichlorosilan
CCl3F
Fluorotrichloromethan
SO2
Sulphur dioxide
CClF3
Chlorotrifluoromethan
WF6
Tungsten hexafluoride
CF4
Carbon tetrafluoride
Xe
Xenon
CH3Br
Bromomethan
The toggles are working. Please, note, that you have to add a template to the library in order to be able to display it inside the toggles.

Electronic pressure regulators are often used to precisely control the pressure in a process chamber or to regulate the pressure in your system. Such a pressure regulator controls a control valve, allowing the pressure in the process chamber to be increased. To reduce the pressure again, vent valves are often used to ensure continuous venting to the environment.

However, this solution is not recommended for expensive or hazardous process gases. In this case, an additional valve with a valve control unit is often used to regulate the venting. These additional components make the system very complex. The process pressure regulator of the P-800 series combines both functions, i.e. it has an integrated venting valve and a venting valve, which reduces the number of components and avoids continuous venting into the environment. The advantages at a glance:

The following gases are included:

Conventional solution with venting to the atmosphere
Process pressure regulator
– Higher gas consumption due to constant venting
– Not suitable for hazardous gases
– Low gas consumption
– Safe solution for all gases
Conventional solution with separate relief valve
Process pressure regulator
– Connection of different components requires more space
– Purchase, assembly and testing of different components is less efficient and more expensive
– Compact, integrated solution
– Economical “plug-and-play” solution
Questions about testing the devices

Calibration

If you click on the link below, you can open an example calibration certificate with explanations of the different sections.

Pdf of calibration certificate

All process instruments are subject to wear and tear due to the process conditions to which they are exposed. Temperature, changes in electronic component tolerance, contamination that accumulates over time and other factors affect accuracy. Your instruments should therefore undergo at least one calibration check, if not recalibration, on a regular basis.

As all applications are different, Bronkhorst does not provide specific due dates for calibration of its instruments. We recommend an annual calibration of the devices. Based on a company’s application conditions and possibly quality assurance procedures, it is up to each customer to determine when to send an instrument in for recalibration. Properly calibrated instruments are more accurate and reliable, ensuring consistency and helping to improve production yields. On-site testing can reduce downtime and costs.

In general, operation with other media is also possible. However, it is important to ensure that the sealing material used is suitable for the new gas or liquid beforehand. In addition, a conversion factor must be used when using a thermal mass flow meter, as the thermodynamic properties of the various media differ. Another influencing factor is the Kv value of the valve. The use of a different medium can significantly change the flow range for which the instrument was designed.

In contrast to thermal mass flow meters, flow meters based on the Coriolis or ultrasonic measuring principle are independent of the medium and do not require a conversion factor. However, if a liquid flow meter is switched to a different medium, the pressure loss may change.

Yes, Bronkhorst instruments can be adapted. After adaptation, the flow meter is recalibrated. The instrument limits must of course be taken into account. We offer repairs and recalibrations throughout Austria.

It is possible to calibrate an instrument by checking the measured value yourself using an exact reference value. It is also possible to restore the function of instruments if no parts are irreversibly damaged. However, this may affect the accuracy of the instruments.

Please note that once an instrument has been opened, the factory calibration and pressure and leak test are no longer valid. Bronkhorst cannot accept any warranty for instruments that have not been opened by certified service partners such as AcuraSens, but by third parties.

In general, we recommend contacting our technical team for repair, cleaning and calibration of contaminated or defective instruments.

Adjusting the measurement signal means that we can adjust the settings of the measuring device to the values of the reference device used. This is to rule out systematic measurement deviations.

Matching is important if the output signal of the measuring device is not only expected to be repeatable, but also to measure the true value. Instead of adjusting the meter, the end user can also use the deviation listed on the calibration certificate to make an adjustment in the computer.

As-found calibration

An as-found calibration is a calibration that is performed before a repair or adjustment. It is often performed when a flow meter has been returned to a service center. This calibration determines the condition the flow meter is in when it is received. On the other hand, there is the as-left calibration.

As-left calibration

The as-left calibration is a calibration that is carried out after a repair or adjustment. It confirms that repairs or adjustments have been successful.

All instruments we offer are calibrated according to ISO 9001. The standard calibration according to this standard is included in the purchase price. It is possible to perform an additional calibration according to ISO 17025 on request.

The specification limit is the maximum permissible value of a measurement deviation. This can be either a positive or a negative deviation.

Questions about our service

Service

You can reach our service team either by email at service@acurasens.at or, in emergencies, by phone
at +43(0)662 439484-10. We will respond reliably within 24 hours.

General inquiry
Service request

Each customer decides for themselves whether and how often they want to send their appliances in for a service. We are happy to provide individual plans and take care of the process. In the event of problems such as deviations or malfunctions, our technical team will carry out a telephone analysis in advance and then decide whether the appliance needs to be sent in or whether an on-site service is possible.

A service always starts with an analysis and a cleaning of the instruments. We clarify any malfunctions and decide whether and how they can be rectified. In any case, the customer must submit a declaration of decontamination in advance.

A declaration of decontamination is used to document all safety-relevant information for the return of instruments and components. The customer is responsible for the health and safety of everyone who may have contact with a returned device. This responsibility extends from the customer’s own personnel to the shipping company and the entire AcuraSens and Bronkhorst team. Any contamination of the device must be identified, even if it has been removed. You can download the decontamination declaration here:

Download

You will receive a cost estimate based on a device analysis carried out by our technicians before each service. AcuraSens offers its customers a three-year warranty on all components.

Calibration is the process of checking the accuracy of a device. In short, calibration is the comparison of the output to a reference. This regular check is important because devices are subject to wear and tear due to process conditions, which can lead to small deviations over time. For some applications, this regular check is even required by law or guidelines. Regular calibration can prevent malfunctions and demonstrably increase the efficiency of a system.

The calibration itself takes a maximum of one hour per device. In the case of on-site calibration, the actual duration depends on how accessible the instruments are to our technicians. For a calibration at our AcuraSens headquarters in Lenzing, the duration of the postal route must be taken into account. We can calibrate most of the instruments ourselves, we only very rarely have to send instruments to Bronkhorst.

Yes, customers can book us for an on-site calibration. From a certain number of devices, this option is even more efficient and cheaper than dismantling and sending in the devices yourself. What is important: For an on-site calibration, we need a test specimen and a reference device. There must also be an area in which our technicians can install the measuring device.

Yes, we are happy to provide your team with advice and ideas by telephone. We offer this form of support free of charge. Please make an appointment and contact us at service@acurasens.at.

Questions about technical problems

Technology

You can find the type of seal in the following table:

model – series
Sealing surface
EL-FLOW Base
BSPP RS-type based sealing surface
EL-FLOW Prestige
BSPP RS-type based sealing surface
EL-FLOW Select
BSPP RS-type based sealing surface
LOW-dP-FLOW
BSPP RS-type based sealing surface
EL-PRESS
BSPP RS-type based sealing surface
EX-FLOW
BSPP RS-type based sealing surface
IN-FLOW
BSPP RS-type based sealing surface
IN-PRESS
BSPP RS-type based sealing surface
LIQUI-FLOW
BSPP RS-type based sealing surface
MASS-STREAM D-6300
BSPP RP-type based sealing surface
MASS-STREAM D-6400
BSPP type ISO 1179-1 flat sealing surface
MASS-VIEW
BSPP type ISO 1179-1 flat sealing surface
FLEXI-FLOW
BSPP type ISO 1179-1 flat sealing surface
IQ+FLOW
10-32 UNF Class 2B sealing surface
EL-FLOW Metal
no sealing surface – welded or part of the body
EL-PRESS Metal
no sealing surface – welded or part of the body
ES-FLOW
no sealing surface – welded or part of the body
µ-FLOW
no sealing surface – welded or part of the body
(mini) CORI-FLOW
no sealing surface – welded or part of the body
IN-FLOW/EX-FLOW F-107/117
no sealing surface – welded or part of the body
IN-FLOW/EX-FLOW F-106
F-106 no sealing surface – wafer type
IP-Rating

The IP-rating of an instrument consists of the letters IP (abbreviation for “Ingress Protection”) followed by two code numbers and an optional letter. As defined in the international standard IEC 60529, it classifies the degree of protection against the ingress of solid foreign bodies including body parts such as hands and fingers, dust and against accidental contact (the first digit after IP) as well as water (the second digit after IP) in electronic housings.

First digit: Protection against solid foreign bodies

Digit
Protection against solid foreign bodies
0
No protection
1
Protected against solid foreign bodies with a diameter of 50 mm or more, access with the back of the hand
2
Protected against solid foreign bodies with a diameter of 12.5 mm or more, access with one finger
3
Protected against solid foreign bodies with a diameter of 2.5 mm or more, access with a tool
4
Protected against solid foreign bodies with a diameter of 1.0 mm or more, access with a wire
5
Protected against dust in damaging quantities, complete protection against contact
6
Dust-tight, complete protection against contact

Second digit: Protection against liquids

Digit
Protection against solid foreign bodies
0
No protection
1
Protection against vertically falling dripping water
2
Protection against dripping water if the housing is inclined up to 15°
3
Protection against falling spray water up to 60° against the vertical
4
Protection against splashing water on all sides
5
Protection against water jets (nozzle) from any angle
6
Protection against strong water jets
7
Protection against temporary immersion between 15 cm and 1 m
8
Protection against permanent immersion
NEMA-Rating

The US National Electrical Manufacturers Association (NEMA) also publishes protection values for enclosures that are comparable to the IP protection system of the International Electrotechnical Commission (IEC). However, NEMA also prescribes other product characteristics that are not covered by the IP codes, such as corrosion resistance, the ageing of seals and construction techniques.

Therefore, while it is possible to equate IP codes to NEMA classifications that meet or exceed the IP code criteria, it is not possible to equate NEMA protection classes to IP codes because the IP codes do not provide for the additional requirements.

IP-Code
min. NEMA protection classes of enclosures to comply with the IP Code
IP20
1
IP54
3
IP65
4, 4X
IP67
6
IP68
6P

The standard material used in the manufacture of all Bronkhorst instruments is 316 stainless steel or equivalent. On request we can also offer some wetted parts in Monel or Hastelloy.

The instruments of the MASS-STREAM, FLEXI-FLOW and IQ+FLOW series are also available in aluminum. Standard seals for thermal mass flow meters and controllers for gases are made of Viton® (FKM) and for thermal mass flow meters and controllers for liquids of Kalrez® (FFKM). (mini) CORI-FLOW Coriolis mass flow meters are metal sealed.

The regulators include internal seals made of Viton® (FKM) (factory standard), EPDM or Kalrez® (FFKM). The sealing materials can also be supplied with FDA certification.

The choice of suitable elastomer seals must be based on chemical resistance under the relevant operating conditions. We are happy to advise our customers on the selection of suitable gaskets. However, our recommendations are to be regarded as guidelines for which no guarantees can be given.

Bronkhorst uses compression or vacuum fittings (VCR/VCO) with BSPP threads as standard for its instruments. BSPP (British Standard Pipe Parallel) is a parallel fitting. The screw threads are used to hold the two pieces together and not to create a seal.

Sealing is achieved using an elastomer seal outside the thread. As the seal is made outside the thread, there is no risk of foreign bodies in the threads being forced into the gas flow path. For this reason, we do not recommend NPT fittings with PTFE sealing tape.